CARELINK
Report
- Report Number
- 2182208-2013-01040
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Report Date
- February 26, 2013
- Manufacturer
- RICE CREEK MFG
- Product Code
- DXY
- PMA / PMN Number
- P890003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS REPORT IS BASED SOLELY ON DEVICE ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THE UPLINK AMPLITUDE WAS OUT OF SPECIFICATION; RF (RADIO FREQUENCY) HEAD CABLE OUT OF ELECTRICAL SPECIFICATION. THE LABEL ALSO NEEDED TO BE REPLACED. (B)(4).
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: ANALYSIS FOUND THE UPLINK AMPLITUDE IS OUT OF SPECIFICATION; RF (RADIO FREQUENCY) HEAD CABLE OUT OF ELECTRICAL SPECIFICATION. (B)(4).
THE RF (RADIO FREQUENCY) HEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
THE RF (RADIO FREQUENCY) HEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT INVOLVEMENT OR COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149766 | CARELINK | PULSE-GENERATOR, PACEMAKER, IMPLANTABLE | DXY | RICE CREEK MFG | 2067 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |