FDA Adverse Event Summary report: N

RELIANCE SYNERGY WASHER

MDR report key: 3050934 · Received April 10, 2013

Report

Report Number
9680353-2013-00035
Date Received
April 10, 2013
Date of Event
March 13, 2013
Report Date
April 10, 2013
Manufacturer
STERIS CANADA CORPORATION
Product Code
MEC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A STERIS SERVICE TECHNICIAN ARRIVED ON SITE, INSPECTED THE UNIT AND FOUND THE DRYING PISTON VALVE BROKEN. THE TECHNICIAN REPLACED THE DRYING PISTON VALVE IN ADDITION TO THE ASSOCIATED HOSE AND SILICONE AND SEAL O-RING. THE TECHNICIAN RAN A TEST CYCLE AND FOUND THE UNIT OPERATIONAL; THE WASHER WAS RETURNED TO SERVICE.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT THEIR RELIANCE SYNERGY WASHER PUMP IS OVERFLOWING AND LEAKING WATER ONTO THE FLOOR. NO PROCEDURAL DELAYS OR CANCELLATIONS OCCURRED. NO INJURIES TO HOSPITAL STAFF OR PATIENTS WERE REPORTED.