RELIANCE SYNERGY WASHER
MDR report key: 3050934
·
Received April 10, 2013
Report
- Report Number
- 9680353-2013-00035
- Date Received
- April 10, 2013
- Date of Event
- March 13, 2013
- Report Date
- April 10, 2013
- Manufacturer
- STERIS CANADA CORPORATION
- Product Code
- MEC
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A STERIS SERVICE TECHNICIAN ARRIVED ON SITE, INSPECTED THE UNIT AND FOUND THE DRYING PISTON VALVE BROKEN. THE TECHNICIAN REPLACED THE DRYING PISTON VALVE IN ADDITION TO THE ASSOCIATED HOSE AND SILICONE AND SEAL O-RING. THE TECHNICIAN RAN A TEST CYCLE AND FOUND THE UNIT OPERATIONAL; THE WASHER WAS RETURNED TO SERVICE.
Description of Event or Problem · 1
THE USER FACILITY REPORTED THAT THEIR RELIANCE SYNERGY WASHER PUMP IS OVERFLOWING AND LEAKING WATER ONTO THE FLOOR. NO PROCEDURAL DELAYS OR CANCELLATIONS OCCURRED. NO INJURIES TO HOSPITAL STAFF OR PATIENTS WERE REPORTED.