FDA Adverse Event Injury Summary report: N

ADAPTA

MDR report key: 3050916 · Received April 10, 2013

Report

Report Number
3004209178-2013-05885
Event Type
Injury
Date Received
April 10, 2013
Report Date
January 2, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
NVZ
PMA / PMN Number
P980035
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT PRODUCTS: 5054 IMPLANTABLE PACING LEAD (B)(6) 2008, 5568 IMPLANTABLE PACING LEAD (B)(6) 2008. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) REACHED ELECTIVE REPLACEMENT INDICATOR EARLY. ALSO REPORTED WAS CHRONIC HIGH THRESHOLD OF THE RIGHT VENTRICULAR (RV) LEAD WITH EXIT BLOCK. THE DEVICE AND THE RV LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148619 ADAPTA PULSE GENERATOR, PERMANENT, IMPLANTABLE NVZ MEDTRONIC MED REL MEDTRONIC PUERTO RICO ADDR01

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| R