FDA Adverse Event
Injury
Summary report: N
ADAPTA
MDR report key: 3050916
·
Received April 10, 2013
Report
- Report Number
- 3004209178-2013-05885
- Event Type
- Injury
- Date Received
- April 10, 2013
- Report Date
- January 2, 2013
- Manufacturer
- MEDTRONIC MED REL MEDTRONIC PUERTO RICO
- Product Code
- NVZ
- PMA / PMN Number
- P980035
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PRODUCT EVENT SUMMARY: THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION. CONCOMITANT PRODUCTS: 5054 IMPLANTABLE PACING LEAD (B)(6) 2008, 5568 IMPLANTABLE PACING LEAD (B)(6) 2008. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) REACHED ELECTIVE REPLACEMENT INDICATOR EARLY. ALSO REPORTED WAS CHRONIC HIGH THRESHOLD OF THE RIGHT VENTRICULAR (RV) LEAD WITH EXIT BLOCK. THE DEVICE AND THE RV LEAD WERE EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148619 | ADAPTA | PULSE GENERATOR, PERMANENT, IMPLANTABLE | NVZ | MEDTRONIC MED REL MEDTRONIC PUERTO RICO | ADDR01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |