FDA Adverse Event Malfunction Summary report: N

LIGACLIP** ENDOSCOPIC CLIP APPLIER

MDR report key: 3050734 · Received April 10, 2013

Report

Report Number
3005075853-2013-01699
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 15, 2013
Report Date
March 18, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION WITH THE JAWS PROPERLY ALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? FIRST. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? CYSTIC DUCT. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? ASKU. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? NO. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD? NO. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? UNKNOWN. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? ASKU.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE WAS FIRED ACROSS THE CYSTIC DUCT AND 3-4 CLIPS THEN NOTICED SCISSORING. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152784 LIGACLIP** ENDOSCOPIC CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1