ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-03990
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 15, 2013
- Report Date
- March 18, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW UP #1 SUBMITTED: 06/06/2013, DEVICE EVALUATION: ON EXAMINATION, THE KEYPAD WAS NOTED TO BE TORN ON THE LEFT SIDE, NEAR THE UP ARROW, DOWN ARROW AND OK KEYPAD BUTTONS. A DAMAGED KEYPAD RUBBER WILL PERMIT CONTAMINATION TO PERMEATE THE BUTTONS WHICH WILL HAVE A NEGATIVE IMPACT ON BUTTON FUNCTION. THIS SITUATION IS NOT LIKELY TO RESULT IN AN ADVERSE EVENT AS THE DAMAGED KEYPAD RUBBER SHOULD BE CLEARLY VISIBLE AND PROHIBIT THE USE OF THE KEYPAD BUTTONS. USERS MAY EXPECT KEYPAD DAMAGE DURING NORMAL WEAR AND THE OWNER¿S BOOKLET INSTRUCTS THE USER TO CONTACT CUSTOMER SERVICE IF THE USER SUSPECTS THE PUMP MAY BE DAMAGED. ON TESTING, THE DOWN ARROW BUTTON HAS INTERMITTENT RESPONSE. THE REMAINDER OF THE BUTTONS HAD NORMAL RESPONSE. THE KEYPAD COVER WAS REMOVED FOR INVESTIGATION AND REVEALED CONTAMINATION UNDER THE CONTACTS OF ALL THE BUTTONS.
ON (B)(6) 2013, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE WITH THE UP/DOWN/OK BUTTONS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 153486 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |