FDA Adverse Event Malfunction Summary report: N

ENSEAL G2 STRAIGHT JAW

MDR report key: 3050676 · Received April 10, 2013

Report

Report Number
3005075853-2013-01695
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 21, 2013
Report Date
March 26, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GEI
PMA / PMN Number
K112033
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED WITH THE TOP OF THE I-BLADE FRACTURED AND LIFTED. IN ADDITION THE JAW WAS FOUND DETACH FROM INSTRUMENT AND NOT RETURNED. THE DEVICE WAS PARTIALLY TESTED WITH A GENERATOR AND THE DEVICE PERFORMED AS REQUIRED DURING TESTING; THOUGH ALL TESTING COULD NOT BE COMPLETED DUE TO THE DAMAGED TOP JAW AND I BLADE. AS REPORTED IN THE EVENT DESCRIPTION IT IS LIKELY THAT THE DAMAGE TO THE JAW AND I-BLADE WAS CAUSED BY NOT ALLOWING THICK AND FIBROUS TISSUES TO DENATURE PRIOR TO I-BLADE ADVANCEMENT OR ATTEMPTING TO CUT THROUGH THICK DENSE TISSUE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC HYSTERECTOMY PROCEDURE THE GYNECOLOGIST TRIED TO DISSECT/CUT A PORTION OF THE WALL OF THE CERVIX. THE TISSUE WAS IN FACT VERY THICK/DENSE AND WHILE CUTTING AND COAGULATING THE JAW OF THE INSTRUMENT BROKE OFF. THE DOCTOR ADMITTED THE CAUSE WAS LIKELY DUE TO THE FACT THAT THE TISSUE WAS PROBABLY TOO THICK BUT SHE WANTED TO SEE THE LIMITS OF THE INSTRUMENT. WE WANT TO REPORT THIS BECAUSE A TOTAL BREAKAGE OF THE JAWS IS NOT VERY NORMAL. NO PATIENT CONSEQUENCES. PROCEDURE WAS COMPLETED WITH USE OF A MONOPOLAR DEVICE. ONE DEVICE WILL BE RETURNING

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153485 ENSEAL G2 STRAIGHT JAW ELECTROSURGICAL CUTTING & COAGULATION & ACCESSORIES GEI ETHICON ENDO-SURGERY, LLC. UNK K4C64N

Patients

Seq Age Sex Outcome Treatment
1 GENERATOR