FDA Adverse Event Malfunction Summary report: N

GEMSTR 7 THPY ALT ST

MDR report key: 3050629 · Received March 13, 2013

Report

Report Number
9615050-2013-00371
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
January 1, 2013
Report Date
February 4, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K060806
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING TESTING, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THIS WAS DUE TO CORROSION ON THE MIDDLE PRINTED CIRCUIT BOARD. THE CAUSE OF THE CORROSION WAS SPILLAGE FROM USE IN THE CUSTOMER ENVIRONMENT. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

DURING VERIFICATION TESTING AT THE SVC CENTER, THE DEVICE DID NOT SOUND AN ALARM TONE DURING AN ALARM CONDITION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106606 GEMSTR 7 THPY ALT ST 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA