FDA Adverse Event
Malfunction
Summary report: N
GEMSTR 7 THPY ALT ST
MDR report key: 3050629
·
Received March 13, 2013
Report
- Report Number
- 9615050-2013-00371
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- January 1, 2013
- Report Date
- February 4, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K060806
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DURING TESTING, THE DEVICE DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THIS WAS DUE TO CORROSION ON THE MIDDLE PRINTED CIRCUIT BOARD. THE CAUSE OF THE CORROSION WAS SPILLAGE FROM USE IN THE CUSTOMER ENVIRONMENT. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
Description of Event or Problem · 1
DURING VERIFICATION TESTING AT THE SVC CENTER, THE DEVICE DID NOT SOUND AN ALARM TONE DURING AN ALARM CONDITION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106606 | GEMSTR 7 THPY ALT ST | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |