SYMBIQ SINGLE CHANNEL
Report
- Report Number
- 9615050-2013-00359
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- January 23, 2013
- Report Date
- February 11, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K041550
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED A WHITESCREEN ERROR. THE PUMP WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "WHITESCREEN ERROR." NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PT INFORMATION, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY AFTER THE DEVICE WAS POWERED ON, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106678 | SYMBIQ SINGLE CHANNEL | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |