FDA Adverse Event Malfunction Summary report: N

NC QUANTUM APEX¿

MDR report key: 3050542 · Received April 10, 2013

Report

Report Number
2134265-2013-02111
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 12, 2013
Report Date
March 12, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LOX
PMA / PMN Number
P860019
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATED BY MANUFACTURER: IT WAS INDICATED THAT THE DEVICE WOULD NOT BE RETURNED FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE LESION BEING TREATED WAS LOCATED IN THE SEVERELY CALCIFIED RIGHT CORONARY ARTERY (RCA). THE PHYSICIAN PERFORMED ROTATION ATHERECTOMY WITH 1.25 AND A 1.75 ROTABLATOR BURRS, THEN ADVANCED A 20MM X 3.00MM NC QUANTUM APEX BALLOON CATHETER TO THE TARGET LESION. THE BALLOON WAS INFLATED AT 14ATMS AND RUPTURED. THE DEVICE WAS SUCCESSFULLY REMOVED AND THE PROCEDURE WAS COMPLETED BY IMPLANTING SIX PROMUS ELEMENT PLUS STENTS IN THE RCA. NO COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS FINE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148642 NC QUANTUM APEX¿ CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS LOX BOSTON SCIENTIFIC - MAPLE GROVE H7493912420300 15468945

Patients

Seq Age Sex Outcome Treatment
1 68 YR