FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 3050540 · Received March 11, 2013

Report

Report Number
2027969-2013-00205
Event Type
Malfunction
Date Received
March 11, 2013
Date of Event
March 6, 2013
Report Date
March 11, 2013
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K092987
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION PENDING.

Description of Event or Problem · 1

"CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS:" - INRATIO 2.8, RETEST 2.3, RETEST 1.3. THERAPEUTIC RANGE: 2-3. TIME BETWEEN TESTS: 5 MINUTES. TESTING WAS PERFORMED USING THE SAME FINGER; MILKING FINGER AFTER FINGER STICK. THE METER WAS NOT IN THE CORRECT MODE WHEN FINGER STICK WAS PERFORMED AND IT TOOK LONGER THAN 15 SECONDS TO APPLY THE SAMPLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
102514 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100071 299622

Patients

Seq Age Sex Outcome Treatment
1 ORAL ANTICOAGULANTS COUMADIN