ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-03971
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Report Date
- March 14, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/14/2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THERE WAS A DIM PINK/FADED DISPLAY SCREEN. THE PUMP WAS OPENED TO REMOVE DISPLAY SCREEN. EVALUATION WAS COMPLETED WITH A TEST DISPLAY SCREEN, THE DISPLAY SCREEN IS SHOWING A STRONG CONTRAST. THE DISPLAY LENS WAS FOUND TO BE SCRATCHED. (B)(6). DEVICE HISTORY RECORD REVIEW WAS CONDUCTED ON 02/12/2013 WITH THE FOLLOWING FINDINGS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE DISPLAY SCREEN BOOTED WITH A DIM, PINK/FADED DISPLAY. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148555 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |