FDA Adverse Event Malfunction Summary report: N

MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT

MDR report key: 3050516 · Received April 10, 2013

Report

Report Number
2955842-2013-01178
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 11, 2013
Report Date
March 11, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THAT ONE OF THE GRIP CABLES WAS BROKEN AT THE DISTAL IDLERS PULLEY AND WAS STICKING OUT AT THE INSTRUMENT'S WRIST. THE IDLER PULLEY SPUN FREELY BUT HAD DAMAGES ON THE SURFACE AND EDGE. NO OTHER CABLES AT THE INSTRUMENT'S WRIST WERE DAMAGED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

THE USER FACILITY REPORTEDLY IDENTIFIED A BROKEN CABLE FROM THE MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT AFTER IT WAS CLEANED AND STERILIZED. THE INSTRUMENT REPORTEDLY WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND HERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
152402 MEGA SUTURECUT NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420309-01 M10120814 377

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES