FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 3050158 · Received April 10, 2013

Report

Report Number
2124215-2013-03906
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
February 18, 2013
Report Date
June 5, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT WAS RECENTLY SEEN POST MAZE PROCEDURE AND IT WAS NOTED THAT THE MAGNET RESPONSE WAS AT 90 WHEN LAST IT WAS AT 100, ALSO THE REMAINING LONGEVITY SHOWED LESS THAN 6 MONTHS REMAINING WHICH WAS LESS THAN EXPECTED BY THE PHYSICIAN. THE PHYSICIAN ALSO STATED THAT THE PATIENT HAS INCREASE ATRIAL OUTPUTS. THE DEVICE REMAINS IN SERVICE. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149076 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1294

Patients

Seq Age Sex Outcome Treatment
1 51 YR 1388T| 1294| 4087| 4088