FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3050104 · Received April 10, 2013

Report

Report Number
2124215-2013-03574
Event Type
Injury
Date Received
April 10, 2013
Date of Event
January 23, 2013
Report Date
June 10, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME, THIS LEAD HAS NOT BEEN RETURNED. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE ON THE HIGH VOLTAGE CHANNEL WHICH WAS RECREATED WHEN THE PATIENT WAS ADJUSTING HIMSELF IN HIS CHAIR. THE IMPEDANCE MEASUREMENTS WERE WITHIN NORMAL LIMITS. IT WAS NOTED THAT THE RATE\SENSE CHANNEL DID NOT HAVE NOISE. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
150184 ENDOTAK RELIANCE NVY CPI - DEL CARIBE 0296

Patients

Seq Age Sex Outcome Treatment
1 40 YR Hospitalization| L| R E160| 0296