ACUITY
Report
- Report Number
- 2124215-2013-03928
- Event Type
- Injury
- Date Received
- April 10, 2013
- Date of Event
- February 20, 2013
- Report Date
- March 28, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. LABORATORY ANALYSIS DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.
THE LEAD WAS SUCCESSFULLY EXPLANTED AND REPLACED. THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE IN-CLINIC FOLLOW UP, THIS LEFT VENTRICULAR (LV) LEAD EXHIBITED LOSS OF CAPTURE (LOC) AT MAXIMUM OUTPUTS. ALL OTHER LEAD DIAGNOSTICS WERE OBSERVED TO BE WITHIN NORMAL LIMITS. X-RAY IMAGING REVEALED THAT THE LEAD HAD DISLODGED. OF NOTE, THE PATIENT REPORTED HANDLING SEVERAL HEAVY ITEMS THE WEEK PRIOR. AN INVASIVE PROCEDURE WAS PERFORMED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148284 | ACUITY | IMPLANTABLE HF LEAD | OJX | CPI - DEL CARIBE | 4592 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | Hospitalization| L| R | 4592| 4469| 0184| T125| MISMATCH| N161 |