FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3050060 · Received April 10, 2013

Report

Report Number
2124215-2013-03535
Event Type
Injury
Date Received
April 10, 2013
Date of Event
February 16, 2013
Report Date
February 27, 2013
Manufacturer
GUIDANT PUERTO RICO BV
Product Code
NVY
PMA / PMN Number
P910073
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE SAVE TO DISK BY ENGINEERING AND TECHNICAL SERVICES NOTED THAT THE DEVICE HAD NO RESETS AND ONE FAULT FOR LOW R WAVE MEASUREMENTS. THE BATTERY STATUS WAS AT BOL, THE DEVICE APPEARED TO BE FUNCTIONING PROPERLY.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION RECEIVED FROM TECHNICAL SERVICES NOTED THAT A RIGHT VENTRICULAR LEAD ISSUE WAS SUSPECTED. TECHNICAL SERVICES DISCUSSED THAT THE PACING INHIBITION NOTED IS MOST LIKELY DUE TO A LEAD INTERMITTENT MECHANICAL ISSUE OR A LEAD/TISSUE INTERFACE PROBLEM OR AN INTERMITTENT CAPTURE WITH THE PREVIOUSLY PROGRAMMED LOWER RIGHT VENTRICULAR OUTPUTS, ALTHOUGH NONE OF THE SCENARIOS CAN BE CONFIRMED AT THIS TIME. AT THIS TIME THE SYSTEM REMAINS IMPLANTED.

Additional Manufacturer Narrative · 1

UPON ADDITIONAL INFORMATION THIS INVESTIGATION WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE SHOWED 10 SECONDS OF ASYSTOLE DUE TO LOSS OF CAPTURE WHICH RESULTED IN SYNCOPE. THE DEVICE RECORDED SEVERAL RVP WHICH WERE NOT FOLLOWED BY RV CAPTURE CAUSING THE PATIENT SYNCOPE. NO NOISE WAS RECORDED AND THE RIGHT VENTRICULAR PACING IMPEDANCES AND PACING THRESHOLDS APPEAR STABLE. TECHNICAL SERVICES WAS CONTACTED, AND DISCUSSED PERFORMING DEEP ISOMETRIC AND POCKET MANIPULATION. A SAVE TO DISK AND MEMORY DUMP WAS REQUESTED FOR FURTHER EVALUATION. AT THIS TIME THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
149556 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY GUIDANT PUERTO RICO BV 0147

Patients

Seq Age Sex Outcome Treatment
1 Life Threatening