FDA Adverse Event Malfunction Summary report: N

INCEPTA

MDR report key: 3050056 · Received April 10, 2013

Report

Report Number
2124215-2013-05506
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
February 6, 2013
Report Date
March 23, 2016
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. LABORATORY ANALYSIS DID NOT IDENTIFY ANY DEVICE CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT'S HOME MONITORING EQUIPMENT DETECTED A LOW AND HIGH OUT OF RANGE SHOCKING IMPEDANCE FOR THIS PATIENT'S RIGHT VENTRICULAR (RV) LEAD. TO DATE, NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION RECEIVED INDICATES THIS DEVICE WAS EXPLANTED DUE TO AN UNRELATED PRODUCT PERFORMANCE ISSUE WHICH HAS BEEN REPORTED IN REPORT # 2124215-2016-03720. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153393 INCEPTA IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E163

Patients

Seq Age Sex Outcome Treatment
1 68 YR T167| E163| 4469| 1861