FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 3050051 · Received April 10, 2013

Report

Report Number
2124215-2013-03940
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
February 14, 2013
Report Date
February 14, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CONSECUTIVE CHARGE TIMES GREATER THAN THE EXTENDED MID-LIFE ERI CHARGE TIME LIMIT. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, ERI WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE.

Description of Event or Problem · 1

THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION AND RETURNED FOR DISPOSAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151985 VITALITY 2 IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T177

Patients

Seq Age Sex Outcome Treatment
1 66 YR 0184| E141| T177