FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3050027 · Received April 10, 2013

Report

Report Number
2531779-2013-03955
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 12, 2013
Report Date
March 13, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 03/23/2015. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/10/2015 WITH THE FOLLOWING FINDINGS:THE BATTERY WAS NOT RETURNED WITH THE PUMP FOR INVESTIGATION. THE ALLEGED ISSUE COULD NOT BE VERIFIED OR DUPLICATED ON INVESTIGATION. THERE WAS EVIDENCE OF MOISTURE DAMAGE FOUND IN THE BATTERY COMPARTMENT. LEAK TESTING REVEALED A BATTERY COMPARTMENT LEAK. THE PUMP WAS OPENED AND THERE WAS EVIDENCE OF MOISTURE DAMAGE INSIDE THE PUMP. UNRELATED TO THE COMPLAINT, INVESTIGATION REVEALED THAT THE DISPLAY WAS DIM AND DISCOLORED.

Description of Event or Problem · 1

ON (B)(6) 2013 THE DISTRIBUTOR CONTACTED ANIMAS ALLEGING THAT THE BATTERY HAD EXPLODED IN THE PUMP. THE DISTRIBUTOR STATED THAT THE PATIENT REMOVED THE BATTERY CAP FOR A ROUTINE BATTERY CHANGE AND FOUND THAT THE TOP OF THE BATTERY HAD SEPARATED FROM THE BODY OF THE BATTERY. UPON INSPECTING THE PUMP, A CRACK WAS FOUND IN THE BATTERY COMPARTMENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
151990 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1