FDA Adverse Event Summary report: N

PTS PANELS GLUCOSE TEST STRIPS

MDR report key: 3049323 · Received April 9, 2013

Report

Report Number
1836135-2013-00001
Date Received
April 9, 2013
Date of Event
March 8, 2013
Report Date
April 4, 2013
Manufacturer
POLYMER TECHNOLOGY SYSTEMS
Product Code
CGA
PMA / PMN Number
010368
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE COMPLAINT WAS RECEIVED FROM (B)(6), WHERE THEY REPORTED UNDER-RECOVERY OF GLUCOSE LOT NUMBER U117 (EXP: 2013 05 08). THE CUSTOMER REPORTED COMPARATIVE DATA BETWEEN GLUCOSE TEST STRIPS LOT NUMBERS U117 AND U204 GLUCOSE RESULTS USING A CARDIOCHEK PA (CCPA). A TOTAL OF THREE TESTS WERE PERFORMED BY THE CUSTOMER WITH GLUCOSE LOT NUMBER U117 STRIPS AND ALL TESTS DEMONSTRATED LOWER THAN EXPECTED RESULTS. A TOTAL OF TWO TESTS WERE PERFORMED BY THE CUSTOMER WITH GLUCOSE LOT NUMBER U204 STRIPS AND ALL RESULTS WERE WITHIN THE ACCEPTABLE LIMITS. THE CUSTOMER RETURNED PRODUCT HAD NOT YET BEEN RETURNED AT THE TIME THIS PIR WAS WRITTEN. UNDER-RECOVERY WAS VERIFIED IN THE QC RETAINED STRIPS FOR GLUCOSE LOT NUMBER U117.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
147998 PTS PANELS GLUCOSE TEST STRIPS NONE CGA POLYMER TECHNOLOGY SYSTEMS U117

Patients

Seq Age Sex Outcome Treatment
1