VENTED AUTOFEED CHAMBER
Report
- Report Number
- 9611451-2013-00255
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 7, 2013
- Report Date
- March 11, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K934140
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). LOT NUMBERS OF RETURNED MR290 AUTOFEED HUMIDIFICATION CHAMBERS: DEVICE 1: LOT 110912, DEVICE 2: LOT 120109, DEVICE 3: LOT 120502, DEVICE 4: LOT 120110. METHOD: FOUR MR290 AUTOFEED HUMIDIFICATION CHAMBERS WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) AND WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A BREAK IN THE WATER FEEDSET TUBES AT THE CONNECTION TO THE CHAMBER DOME ON ALL FOUR MR290 AUTOFEED HUMIDIFICATION CHAMBERS. THE SURFACE OF THE BREAK WAS ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED TWO OTHER COMPLAINTS OF THIS NATURE ONE FROM LOT 120109 AND ONE FROM LOT 120502. CONCLUSION: THE FEEDSET TUBE BREAK WAS ROUGH ON ALL FOUR MR290 AUTOFEED HUMIDIFICATION CHAMBERS, SUGGESTING THAT THE DAMAGE MAY HAVE OCCURRED AS A RESULT OF THE TUBE BEING PULLED AWAY FROM THE CHAMBER POSSIBLY DUE TO THE FEEDSET BEING CAUGHT OR UNDER TENSION. ALL CHAMBERS ARE PRESSURE TESTED BEFORE THEY LEAVE THE PRODUCTION LINE AND ANY HOLES OR LEAKS IN THE FEEDSET ARE IDENTIFIED DURING THIS PROCESS. THIS SUGGESTS THAT THE DAMAGE OCCURRED POST PRODUCTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: "SET APPROPRIATE VENTILATOR ALARMS." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO OUR DISTRIBUTOR THAT THE HEATER WIRE ADAPTOR WAS FAULTY ON AN RT200 ADULT BREATHING CIRCUIT. UPON INSPECTING THE RETURNED COMPLAINT DEVICES THE WATER FEEDSET TUBES OF THE MR290 AUTOFEED HUMIDIFICATION CHAMBERS INCLUDED IN THE RT200 BREATHING CIRCUIT KIT WERE FOUND TO BE DAMAGED. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149647 | VENTED AUTOFEED CHAMBER | AUTOFEED CHAMBER | BTT | FISHER & PAYKEL HEALTHCARE LTD | MR290V | 110912 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | RT200 ADULT BREATHING CIRCUIT |