FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3043538 · Received April 9, 2013

Report

Report Number
2531779-2013-03855
Event Type
Malfunction
Date Received
April 9, 2013
Report Date
March 14, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 06/25/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/29/2013 WITH THE FOLLOWING FINDINGS: THE PUMP¿S HISTORY SHOWED NO EVIDENCE OF REBOOTS. DURING EVALUATION, NO DAMAGE WAS OBSERVED TO THE PUMP¿S BATTERY COMPARTMENT. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP, AND A TEST CAP SECURED ON TIGHTLY WITH NO ISSUES. THE PUMP WAS EXERCISED FOR 24 HOURS WITH A 1 UNIT PER HOUR BASAL PROGRAM, NO POWER INTERRUPTIONS, ERRORS, ALARMS, OR WARNINGS OCCURRED. THE PUMP COVER WAS REMOVED AND NO INTERNAL MOISTURE OR DAMAGE WAS FOUND TO THE POWER CIRCUIT OR BATTERY FLEX.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THE REPORTER INDICATED THAT ON ONE OCCASION THE PUMP SCREEN LIT UP WITHOUT USER PRESSES AND THEN THE PUMP WENT DEAD WITH NO AUDIO TONES OR VIBRATION AND A BLANK SCREEN. THE REPORTER CONFIRMED NO POWER LOSS ISSUES SINCE THE INITIAL EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146810 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 26 YR