ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-03855
- Event Type
- Malfunction
- Date Received
- April 9, 2013
- Report Date
- March 14, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
FOLLOW-UP # 1 DATE OF SUBMISSION 06/25/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 05/29/2013 WITH THE FOLLOWING FINDINGS: THE PUMP¿S HISTORY SHOWED NO EVIDENCE OF REBOOTS. DURING EVALUATION, NO DAMAGE WAS OBSERVED TO THE PUMP¿S BATTERY COMPARTMENT. THE BATTERY CAP WAS NOT RETURNED WITH THE PUMP, AND A TEST CAP SECURED ON TIGHTLY WITH NO ISSUES. THE PUMP WAS EXERCISED FOR 24 HOURS WITH A 1 UNIT PER HOUR BASAL PROGRAM, NO POWER INTERRUPTIONS, ERRORS, ALARMS, OR WARNINGS OCCURRED. THE PUMP COVER WAS REMOVED AND NO INTERNAL MOISTURE OR DAMAGE WAS FOUND TO THE POWER CIRCUIT OR BATTERY FLEX.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (INTERMITTENT POWER) ISSUE. THE REPORTER INDICATED THAT ON ONE OCCASION THE PUMP SCREEN LIT UP WITHOUT USER PRESSES AND THEN THE PUMP WENT DEAD WITH NO AUDIO TONES OR VIBRATION AND A BLANK SCREEN. THE REPORTER CONFIRMED NO POWER LOSS ISSUES SINCE THE INITIAL EVENT. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 146810 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 26 YR |