FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3043536 · Received April 9, 2013

Report

Report Number
2531779-2013-03856
Event Type
Malfunction
Date Received
April 9, 2013
Report Date
March 11, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON 03/11/2013 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE READING LOW FORCE. THE FORCE SENSOR RESISTANCE WAS FOUND TO BE HIGH. CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME SUCCESSFULLY. UNRELATED TO THE EVENT THE DISPLAY WAS FOUND TO BE FADED AND DISCOLORED.

Description of Event or Problem · 1

THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE READING LOW FORCE. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144817 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 34 YR