FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3043536
·
Received April 9, 2013
Report
- Report Number
- 2531779-2013-03856
- Event Type
- Malfunction
- Date Received
- April 9, 2013
- Report Date
- March 11, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON 03/11/2013 WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE READING LOW FORCE. THE FORCE SENSOR RESISTANCE WAS FOUND TO BE HIGH. CONTAMINATION WAS OBSERVED ON THE FORCE SENSOR ASSEMBLY. DURING EVALUATION THE PUMP WAS ABLE TO REWIND, LOAD AND PRIME SUCCESSFULLY. UNRELATED TO THE EVENT THE DISPLAY WAS FOUND TO BE FADED AND DISCOLORED.
Description of Event or Problem · 1
THE PUMP HAS BEEN RETURNED AND WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013, WITH THE FOLLOWING FINDINGS: THE FORCE SENSOR WAS FOUND TO BE READING LOW FORCE. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 144817 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |