FDA Adverse Event
Malfunction
Summary report: N
SPECTRUM INFUSION PUMP
MDR report key: 3043467
·
Received March 12, 2013
Report
- Report Number
- 1314492-2013-00226
- Event Type
- Malfunction
- Date Received
- March 12, 2013
- Date of Event
- February 9, 2013
- Report Date
- February 12, 2013
- Manufacturer
- BAXTER HEALTHCARE CORP.
- Product Code
- FRN
- PMA / PMN Number
- K042121
- Removal / Correction Number
- 1314492-080712-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS RETURNED AND ELEVATED BY BAXTER. THE DEVICE EVAL CONFIRMED THE #2, #3, #4, #5, #6, #7, #8, #9 AND (.) KEY TO BE INOPERABLE CAUSED BY A FAILED KEYPAD. IT WAS OBSERVED THAT WHEN ANY REMAINING FUNCTIONAL KEYS ARE SELECTED, AN AUTOMATIC OUTPUT OF THE #3 KEY WILL OCCUR FOLLOWING THE SELECTED KEY'S FUNCTION (E.G. WHEN THE #1 KEY IS PRESSED 13 WILL BE DISPLAYED. WHEN IN ALPHABETIC MODE AND THE ABC KEY IS SELECTED, AG WILL BE DISPLAYED INTERFERING WITH THE MDL DRUG SEARCH). THE FAILED KEYPAD WAS REPLACED. BAXTER'S ENGINEERING INVESTIGATION IS IN PROGRESS. WHEN COMPLETE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE #3 KEY ON A PUMP KEYPAD IS "STUCK." IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104506 | SPECTRUM INFUSION PUMP | INFUSION PUMP | FRN | BAXTER HEALTHCARE CORP. | 35700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |