FDA Adverse Event Malfunction Summary report: N

CURLIN INFUSION IV PUMP

MDR report key: 3043431 · Received March 11, 2013

Report

Report Number
1722139-2013-00152
Event Type
Malfunction
Date Received
March 11, 2013
Date of Event
February 13, 2013
Report Date
February 14, 2013
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
FRN
PMA / PMN Number
K981816
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

VISUAL INSPECTION OF THE PLATEN DOOR SHOWS NO PHYSICAL DAMAGE. IT OPENS AND CLOSES AS SPECIFIED. ADMINISTRATION SET WAS INSERTED, PLATEN DOOR WAS CLOSED, AND NO FREE FLOW WAS OBSERVED. PERFORMED 40 PSI TEST AND THE PUMP PASSED THE TEST WITH NO ERRORS OR ALARMS. COMPLAINT COULD NOT BE CONFIRMED.

Description of Event or Problem · 1

CUSTOMER STATES THAT DURING TESTING PUMP CONSTANTLY FREE FLOWS WHEN CASSETTE IS ATTACHED. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
102880 CURLIN INFUSION IV PUMP FRN MOOG MEDICAL DEVICES GROUP 4000 CMS

Patients

Seq Age Sex Outcome Treatment
1