FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3043411 · Received April 9, 2013

Report

Report Number
2531779-2013-03848
Event Type
Malfunction
Date Received
April 9, 2013
Report Date
March 14, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/14/2013 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE DISPLAY SCREEN WAS FOUND TO BE DIM/FADED. UNRELATED TO THE COMPLAINT, TESTING REVEALED THE TIME AND DATE RESET TO FACTORY SETTINGS. THE PUMP WAS OPENED AND EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE PC BOARD HAD FAILED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THE DISPLAY SCREEN WAS DIM/FADED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 03/14/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
147203 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 63 YR