OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2013-00212
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Report Date
- February 11, 2013
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVAL. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. THE CUSTOMER REPORTED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE. THIS CONDITION COULD INTERRUPT INSULIN DELIVERY AND CONTRIBUTE TO HYPERGLYCEMIA. NO LOT QUALIFICATION RECORDS WERE REVIEWED, AS NO PRODUCT LOT NUMBER WAS PROVIDED. THE OMNIPOD USER'S GUIDE WARNS "CHECK OFTEN TO MAKE SURE THE POD AND SOFT CANNULA ARE SECURELY ATTACHED AND IN PLACE. A LOOSE OR DISLODGED CANNULA MAY INTERRUPT INSULIN DELIVERY," AND "BECAUSE INSULIN PODS USE ONLY RAPID-ACTING INSULIN, USERS ARE AT INCREASED RISK FOR DEVELOPING HYPERGLYCEMIA (HIGH BLOOD GLUCOSE) IF INSULIN DELIVERY IS INTERRUPTED. IF IT IS UNTREATED, SEVERE HYPERGLYCEMIA CAN QUICKLY LEAD TO DIABETIC KETOACIDOSIS (DKA). DKA CAN CAUSE BREATHING DIFFICULTIES, SHOCK, COMA, OR DEATH. IF INSULIN DELIVERY IS INTERRUPTED FOR ANY REASON, YOU MAY NEED TO REPLACE THE MISSING INSULIN - USUALLY WITH AN INJECTION OF RAPID-ACTING INSULIN. ASK YOUR HEALTHCARE PROVIDER FOR INSTRUCTIONS ON HANDLING INTERRUPTED INSULIN DELIVERY." IT ALSO WARNS, "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER." IT ADVISES, "TO AVOID HYPERGLYCEMIA (HIGH BLOOD GLUCOSE), CHECK YOUR BLOOD GLUCOSE AT LEAST 4-6 TIMES A DAY (WHEN YOU WAKE UP, BEFORE EACH MEAL, AND BEFORE GOING TO BED)".
THE CUSTOMER REPORTED THAT HER BLOOD GLUCOSE LEVEL REACHED 300 MG/DL TWICE AND THAT SHE WAS NOT FEELING WELL AND WAS "ALMOST IN DKA". SHE SAID THAT HER SKIN WAS WET AND SHE BELIEVES THE CANNULA DISLODGED AND THEN WENT BACK INTO HER SKIN. THE CANNULA WAS ALSO BENT. SHE DOES NOT HAVE THE POD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106258 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |