FDA Adverse Event Malfunction Summary report: N

MARYLAND BIPOLAR FORCEPS INSTRUMENT

MDR report key: 3043367 · Received April 9, 2013

Report

Report Number
2955842-2013-01160
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
February 13, 2013
Report Date
March 13, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THAT THE INSTRUMENT'S TIPS WERE OUT OF ALIGNMENT. ONE OF THE GRIP WAS BENT CAUSING SIDE TO SIDE MISALIGNMENT OF THE GRIPS. THERE WAS A 0.069 OFFSET AT THE TIPS, INDICATING OVERLOADING AT THE TIP. ELECTRICAL CONTINUITY TESTING WAS PERFORMED AND PASSED. ADDITIONAL OBSERVATION NOT INITIALLY REPORTED BY THE SITE WAS MAIN TUBE DAMAGE. ENGINEERING EVALUATION ALSO FOUND THAT THE DISTAL END OF THE INSTRUMENT'S MAIN TUBE EXHIBITED VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL AND HAD A ROUGH SURFACE FINISH. THE SCRATCHES WERE SHORT IN LENGTH AND WERE NOT AXIALLY ALIGNED WITH THE TUBE. ENGINEERING CONCLUDED THAT THE DAMAGE MAY BE DUE TO MISHANDLING. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS - HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO REOCCUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE USER FACILITY IDENTIFIED TIPS ON THE MEGA NEEDLE DRIVER INSTRUMENT THAT WERE ALLEGEDLY OUT OF ALIGNMENT. THE INSTRUMENT REPORTEDLY WAS NOT USED ON A PATIENT AFTER THE REPORTED ISSUE WAS IDENTIFIED AND THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146457 MARYLAND BIPOLAR FORCEPS INSTRUMENT ENDOSCOPIC ELECTROSURGICAL INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420172-07 M10120919 339

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS & ACCESSORIES, ESU