FDA Adverse Event
Malfunction
Summary report: N
E SERIES DEFIBRILLATOR
MDR report key: 3043356
·
Received March 8, 2013
Report
- Report Number
- 1220908-2013-00532
- Event Type
- Malfunction
- Date Received
- March 8, 2013
- Report Date
- February 19, 2013
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- MKJ
- PMA / PMN Number
- K042007
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ZOLL MEDICAL CORPORATION HAS NOT RECEIVED THE PRODUCT FOR EVAL AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.
Description of Event or Problem · 1
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE'S DISPLAY FROZE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 100703 | E SERIES DEFIBRILLATOR | E SERIES | MKJ | ZOLL MEDICAL CORPORATION | E SERIES | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |