ENDOTAK ENDURANCE
Report
- Report Number
- 2124215-2013-02721
- Event Type
- Malfunction
- Date Received
- April 9, 2013
- Date of Event
- January 17, 2013
- Report Date
- January 22, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THAT THIS PATIENT'S HOME MONITORING SYSTEM DETECTED A LOW, OUT-OF-RANGE (OOR), SHOCK LEAD IMPEDANCE FOR THIS RIGHT VENTRICULAR (RV) LEAD AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM. THE FIELD REPRESENTATIVE CALLED TO REVIEW THE INFORMATION WITH BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) AND IT WAS DISCUSSED THAT THE IMPEDANCE MEASUREMENT WAS 0 OHMS AND ALL OTHERS HAVE BEEN IN THE 40-50 OHM RANGE. THE PATIENT WAS SEEN FOR CLINICAL EVALUATION AND IT WAS NOTED THAT THE EVALUATION WAS FINE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE SYSTEM REMAINS IN SERVICE.