ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-01815
- Event Type
- Malfunction
- Date Received
- April 9, 2013
- Date of Event
- January 17, 2013
- Report Date
- February 20, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT TO THE POST MARKET QUALITY ASSURANCE LABORATORY, ANALYSIS WILL BE PERFORMED.
UPON RECEIPT TO THE POST MARKET QUALITY ASSURANCE LABORATORY, THE ENTIRE LEAD WAS RECEIVED FOR ANALYSIS. VISUAL OBSERVATION REVEALED NO ABNORMALITIES AT THE TIP REGION OF THE LEAD. RESISTANCE TESTING WAS COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. ANALYSIS COULD NOT CONFIRM ANY ABNORMALITY OF THE LEAD TIP SECTION OR HELIX THAT MAY HAVE CONTRIBUTED TO THE REPORTED CLINICAL ALLEGATION.
ACCORDING TO AVAILABLE INFORMATION, THIS LEAD AND DEVICE REMAIN IMPLANTED AND IN SERVICE. WHEN ADDITIONAL INFORMATION BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.
ADDITIONAL INFORMATION WAS OBTAINED. A REVISION PROCEDURE WAS PERFORMED. UPON OPENING THE POCKET, NO CONNECTION ISSUE WAS OBSERVED. A DECISION WAS MADE TO REPLACE THIS LEAD. DIFFICULTY WAS ENCOUNTERED IMPLANTING THE REPLACEMENT LEAD DUE TO SENSING ISSUES. FINALLY, AN ACCEPTABLE POSITION WAS FOUND. THIS LEAD WAS REMOVED, REPLACED AND RETURNED FOR ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT APPROXIMATELY ONE MONTH POST IMPLANT, THIS RIGHT VENTRICULAR (RV) LEAD AND CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DISPLAYED INCREASED THRESHOLD MEASUREMENTS THAT WERE INCONSISTENT AND INCREASING. IN ADDITION, THE INCREASED THRESHOLD MEASUREMENTS WERE ALSO OBSERVED IN THE LEFT VENTRICULAR (LV) LEAD. AS THERE WAS LOSS OF BIV PACING, LEFT VENTRICULAR LEAD DISLODGEMENT WAS SUSPECTED. AN INTERNAL TECHNICAL SERVICE CONSULTANT WAS CONTACTED AND RECOMMENDED FURTHER LEAD INTERROGATION AND VERIFICATION OF THE DF-4 LEAD CONNECTION. NO OUT OF RANGE IMPEDANCE MEASUREMENTS HAD BEEN OBSERVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION HAS BEEN RECEIVED. FURTHER REVIEW OF THIS RV LEAD AND DEVICE REVEALED THE DEVICE HAS MIGRATED DOWNWARDS FROM THE IMPLANTED POSITION. A REVIEW OF THE DAILY MEASUREMENTS REVEALED THE SHOCK IMPEDANCE HAS DECREASED SLIGHTLY, HOWEVER REMAINS WITHIN NORMAL LABELING GUIDELINES. THIS LEAD IMPEDANCE HAS DECREASED SIX HUNDRED OHMS SINCE THE LAST MEASUREMENT AND THE RV THRESHOLDS HAVE INCREASED. IT WAS THOUGHT THE DEVICE HAS MIGRATED AND THE LV LEAD IS DISLODGED. A SYSTEM REVISION IS INTENDED AND FURTHER RV LEAD INTEGRITY IS RECOMMENDED. IT WAS THOUGHT THERE MAY BE A CONNECTION ISSUE HOWEVER AS THE RV LEAD IMPEDANCE MEASUREMENTS ARE WITHIN NORMAL LABELING RANGE, THIS MAY SUGGEST A LEAD ISSUE. IT WAS FURTHER THOUGHT THAT AS THE DEVICE HAS MIGRATED, THE RV LEAD MAY HAVE ALSO STRETCHED AND MAY BE CONTRIBUTING TO THE REPORTED CLINICAL OBSERVATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 147772 | ENDOTAK RELIANCE | NVY | CPI - DEL CARIBE | 0292 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |