INGENIO
Report
- Report Number
- 2124215-2013-01381
- Event Type
- Malfunction
- Date Received
- April 9, 2013
- Date of Event
- January 17, 2013
- Report Date
- March 13, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION,PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING AN IMPLANT PROCEDURE THIS DEVICE WAS CONNECTED TO THE EXISTING RIGHT VENTRICULAR (RV) LEAD. THE VENTRICULAR LEAD EXHIBITED HIGH PACING IMPEDANCES GREATER THAN 2,000 OHMS. THE DEVICE WAS DISCONNECTED FROM THE LEAD AND REINSERTED INTO THE LEAD MULTIPLE TIMES TO CHECK FOR A CONNECTION ISSUES. HIGH PACING IMPEDANCES CONTINUED TO OCCUR DURING DEVICE TESTING. THE DEVICE WAS REMOVED AND REPLACED WITH ANOTHER DEVICE. A NEW DEVICE WAS CONNECTED WITH THE EXISTING LEADS;TESTED AND ALL ISSUES WERE RESOLVED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE DEVICE WAS RETURNED FOR ANALYSIS .
---
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 145643 | INGENIO | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | J172 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |