FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 3043116 · Received April 9, 2013

Report

Report Number
2124215-2013-01848
Event Type
Injury
Date Received
April 9, 2013
Date of Event
January 23, 2013
Report Date
January 23, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED AND AN AMENDED REPORT BE SUBMITTED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD WAS SURGICALLY ABANDONED DUE TO A LEAD FRACTURE ISSUE OBSERVED DURING A RECENT DEVICE REPLACEMENT PROCEDURE. DURING THE REPLACEMENT, IT WAS ALSO NOTED THAT THE RV LEAD EXHIBITED HIGH PACING IMPEDANCES GREATER THAN 2,000 OHMS, NOISE AND A LOSS OF CAPTURE (LOC). A NEW LEAD WAS SUCCESSFULLY IMPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
146173 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4457

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L