FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3043075 · Received April 9, 2013

Report

Report Number
1416980-2013-08807
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 17, 2013
Report Date
March 17, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER IS UNKNOWN, A DEVICE ANALYSIS CANNOT BE COMPLETED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BAXTER?S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 ALARM (AIR IN TUBING), WHICH OCCURRED ON THE HOMECHOICE (HC) DURING USE, DURING DRAIN 3 OF 5. THE HP SAW AIR THAT APPEARED TO BE COMING FROM THE TRANSFER SET. THE HP STATED THAT HE HAS AN APPOINTMENT FOR THE TRANSFER SET TO BE REPLACED. THERE WAS PATIENT INVOLVEMENT BUT NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. THE HOME PATIENT WAS CONTACTED AND HE SAID HE WENT TO SEE HIS NURSE THAT DAY AND THEY WERE NOT ABLE TO DETERMINE WHERE THE AIR CAME FROM. HE SAID THERE WAS NO LEAK ON HIS TRANSFER SET OR ANYWHERE ELSE. HE SAID THAT HIS NURSE CHANGED HIS TRANSFER SET THAT DAY AND SINCE THEN HE HAS BEEN COMPLETING THERAPY SUCCESSFULLY. HIS NURSE HAD HIM TAKE ANTIBIOTICS AS A PRECAUTIONARY MEASURE. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
147345 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 69 YR HOME CHOICE