FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3043065
·
Received March 15, 2013
Report
- Report Number
- 3008642652-2013-00709
- Event Type
- Malfunction
- Date Received
- March 15, 2013
- Date of Event
- February 25, 2013
- Report Date
- March 11, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVAL, THE CABLE THAT CONNECTS THE DISTRIBUTION NODE (DN) TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, WHICH CAUSED THE ELECTRODE BELT TO FAIL A HI-POT TEST. THE ROOT CAUSE FOR THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED, BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) FAILED A HI-POT TEST DUE TO DAMAGED CABLES. THE LAST PT TO USE THIS ELECTRODE BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 110330 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |