FDA Adverse Event Malfunction Summary report: N

MEDRAD VERIS MR VITAL SIGNS MONITOR

MDR report key: 3043055 · Received March 7, 2013

Report

Report Number
2520313-2013-00015
Event Type
Malfunction
Date Received
March 7, 2013
Date of Event
August 7, 2012
Report Date
August 7, 2012
Manufacturer
MEDRAD, INC.
Product Code
MWI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) SERVICE CONFIRMED WITH THE CUSTOMER THAT THE POWER SUPPLY CORD WAS EXHIBITING AN ELEVATED TEMPERATURE AND THE VERIS MONITOR WAS SHUTTING OFF AND WOULD NOT CHARGE WITH THE EXISTING CHARGER. A NEW CHARGER AND POWER CORD WERE SENT TO THE CUSTOMER, WHICH REMEDIED THE REPORTED ISSUE. A SUBSEQUENT INVESTIGATION OF SIMILAR COMPLAINTS CONCLUDED THAT WHEN CERTAIN COMBINATIONS OF CONDUCTORS SHORT TOGETHER AT THE CABLE END FARTHEST FROM THE POWER SUPPLY, THIS CAN CREATE A CONDUCTION PATHWAY THAT THE POWER SUPPLY DOES NOT RECOGNIZE AS A SHORT CIRCUIT AND SHUT DOWN. CONSEQUENTLY, THE POWER SUPPLY CONTINUES TO ENERGIZE THE CABLE WHICH CAN LEAD TO LOCALIZED HEATING AND MELTING OF THE POWER CABLE.

Description of Event or Problem · 1

THE SITE REPORTED THE FOLLOWING: THE VERIS POWER SUPPLY CORD IS GETTING HOT, THE VERIS MONITOR SHUT OFF.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
98471 MEDRAD VERIS MR VITAL SIGNS MONITOR MONITOR, PHYSIOLOGICAL PATIENT MWI MEDRAD, INC. 3011996W

Patients

Seq Age Sex Outcome Treatment
1