MEDRAD VERIS MR VITAL SIGNS MONITOR
Report
- Report Number
- 2520313-2013-00015
- Event Type
- Malfunction
- Date Received
- March 7, 2013
- Date of Event
- August 7, 2012
- Report Date
- August 7, 2012
- Manufacturer
- MEDRAD, INC.
- Product Code
- MWI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4) SERVICE CONFIRMED WITH THE CUSTOMER THAT THE POWER SUPPLY CORD WAS EXHIBITING AN ELEVATED TEMPERATURE AND THE VERIS MONITOR WAS SHUTTING OFF AND WOULD NOT CHARGE WITH THE EXISTING CHARGER. A NEW CHARGER AND POWER CORD WERE SENT TO THE CUSTOMER, WHICH REMEDIED THE REPORTED ISSUE. A SUBSEQUENT INVESTIGATION OF SIMILAR COMPLAINTS CONCLUDED THAT WHEN CERTAIN COMBINATIONS OF CONDUCTORS SHORT TOGETHER AT THE CABLE END FARTHEST FROM THE POWER SUPPLY, THIS CAN CREATE A CONDUCTION PATHWAY THAT THE POWER SUPPLY DOES NOT RECOGNIZE AS A SHORT CIRCUIT AND SHUT DOWN. CONSEQUENTLY, THE POWER SUPPLY CONTINUES TO ENERGIZE THE CABLE WHICH CAN LEAD TO LOCALIZED HEATING AND MELTING OF THE POWER CABLE.
THE SITE REPORTED THE FOLLOWING: THE VERIS POWER SUPPLY CORD IS GETTING HOT, THE VERIS MONITOR SHUT OFF.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 98471 | MEDRAD VERIS MR VITAL SIGNS MONITOR | MONITOR, PHYSIOLOGICAL PATIENT | MWI | MEDRAD, INC. | 3011996W |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |