GREENLIGHT MOXY FIBER OPTIC
Report
- Report Number
- 2937094-2013-00318
- Event Type
- Malfunction
- Date Received
- March 6, 2013
- Date of Event
- December 29, 2012
- Report Date
- December 31, 2012
- Manufacturer
- AMERICAN MEDICAL SYSTEMS
- Product Code
- GEX
- PMA / PMN Number
- K120870
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
FIBER ANALYSIS: THE FIBER WAS FOUND TO HAVE A CIRCUMFERENTIAL FRACTURE OF THE GLASS CAP, DISTAL TO THE FIBER/CAP FUSION ZONE; THE FIBER PROXIMAL TO THE FRACTURE WAS FOUND TO ROTATE INDEPENDENTLY OF THE METAL CAP. BURNT ON DETRITUS AND DEVITRIFICATION OF THE CAP OUTPUT WINDOW WAS ALSO OBSERVED. BOTH CAPS REMAIN INTACT AND ATTACHED. THE IDENTIFIED ISSUES MAY ACTIVATE THE FIBERLIFE FUNCTION WHICH WOULD MODULATE THE POWER SHOWING A PULSING BEAM OR THE SYSTEM WOULD BE PLACED INTO STANDBY MODE. THIS ISSUE MAY ALSO CAUSE FORWARD-FIRING. THE PROBABLE ROOT CAUSE OF THE DEVICE FAILURE WAS DETERMINED TO BE HEAT ACCUMULATION DUE TO TISSUE CONTACT AND OR ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE.
IT WAS REPORTED THAT WHILE USING THE SURGICAL FIBER DURING A PROSTATE PROCEDURE, THE SYSTEMS FIBERLIFE FUNCTION CONTINUOUSLY ACTIVATED, SENDING THE SYSTEM TO STANDBY MODE AT 96,860 JOULES OF USE. THE CASE WAS COMPLETED USING A SECOND FIBER. THERE WAS NO REPORT OF INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 96414 | GREENLIGHT MOXY FIBER OPTIC | POWERED SURGICAL LASER INSTRUMENT | GEX | AMERICAN MEDICAL SYSTEMS | 10-2400 | 241A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ACCESSORIES| GREENLIGHT XPS LASER SYSTEMS |