FDA Adverse Event Malfunction Summary report: N

EX-PRESS MINI GLAUCOMA SHUNT

MDR report key: 3042912 · Received March 5, 2013

Report

Report Number
3003701944-2013-00017
Event Type
Malfunction
Date Received
March 5, 2013
Report Date
February 5, 2013
Manufacturer
OPTONOL, LTD
Product Code
KYF
PMA / PMN Number
K030350
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: NO SAMPLE WAS RETURNED, THEREFORE, THE CONDITION OF THE PRODUCT COULD NOT BE VERIFIED AND VISUAL INSPECTION CANNOT BE PERFORMED. THE DEVICE HISTORY RECORD (DHR) FOR THE BATCH WAS REVIEWED. NO ABNORMALITIES WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO RELEASE CRITERIA. BECAUSE A SAMPLE WAS NOT RETURNED, THE ROOT CAUSE CANNOT BE DETERMINED. THERE HAS BEEN ONE OTHER SIMILAR COMPLAINT REPORTED IN THE LOT NUMBER. ADDITIONAL INFO HAS BEEN REQUESTED. (B)(4).

Description of Event or Problem · 1

A SURGEON REPORTED THAT AFTER A GLAUCOMA FILTRATION DEVICE IMPLANTATION, HE OBSERVED THAT IT WAS NOT DRAINING. HE REMOVED THE GLAUCOMA FILTRATION DEVICE AND COMPLETED THE SURGERY WITH ANOTHER ONE. THE PT HAD A GOOD OUTCOME WITHOUT ANY COMPLICATIONS. ADDITIONAL INFO HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
94651 EX-PRESS MINI GLAUCOMA SHUNT AQUEOUS SHUNT KYF OPTONOL, LTD P-50 PL 120513

Patients

Seq Age Sex Outcome Treatment
1