FDA Adverse Event Malfunction Summary report: N

ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM

MDR report key: 3042899 · Received April 4, 2013

Report

Report Number
2028159-2013-00624
Event Type
Malfunction
Date Received
April 4, 2013
Date of Event
March 6, 2013
Report Date
March 8, 2013
Manufacturer
ALCON - IRVINE TECHNOLOGY CENTER
Product Code
MPA
PMA / PMN Number
K951627
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND PERFORMED AN INTERNAL AND EXTERNAL CLEANING OF THE EQUIPMENT. THE COMPANY REPRESENTATIVE CLEANED THE FILTERS AND MIRRORS. THE FAN FILTER WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. IT SHOULD BE NOTED THAT A DIRTY FAN FILTER CAN BLOCK THE AIR PATHWAYS CAUSING THE SYSTEM TO OVERHEAT AND SHUT DOWN. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST (B)(4) DID INDICATE 1 SIMILAR REPORT FOR THIS SYSTEM. THE ROOT CAUSE COULD NOT BE CONCLUSIVELY DETERMINED. HOWEVER, A POTENTIAL ROOT CAUSE FOR THIS REPORTED EVENT WAS DETERMINED TO BE A DIRTY FAN FILTER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE ILLUMINATION SYSTEM TURNED OFF ON ITS OWN DURING A PROCEDURE. FOLLOWING A DELAY OF MORE THAN 15 MINUTES, THE PROCEDURE WAS COMPLETED WITH NO IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139116 ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM ENDOILLUMINATOR MPA ALCON - IRVINE TECHNOLOGY CENTER NA NA

Patients

Seq Age Sex Outcome Treatment
1