ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM
Report
- Report Number
- 2028159-2013-00624
- Event Type
- Malfunction
- Date Received
- April 4, 2013
- Date of Event
- March 6, 2013
- Report Date
- March 8, 2013
- Manufacturer
- ALCON - IRVINE TECHNOLOGY CENTER
- Product Code
- MPA
- PMA / PMN Number
- K951627
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM AND PERFORMED AN INTERNAL AND EXTERNAL CLEANING OF THE EQUIPMENT. THE COMPANY REPRESENTATIVE CLEANED THE FILTERS AND MIRRORS. THE FAN FILTER WAS REPLACED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. IT SHOULD BE NOTED THAT A DIRTY FAN FILTER CAN BLOCK THE AIR PATHWAYS CAUSING THE SYSTEM TO OVERHEAT AND SHUT DOWN. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST (B)(4) DID INDICATE 1 SIMILAR REPORT FOR THIS SYSTEM. THE ROOT CAUSE COULD NOT BE CONCLUSIVELY DETERMINED. HOWEVER, A POTENTIAL ROOT CAUSE FOR THIS REPORTED EVENT WAS DETERMINED TO BE A DIRTY FAN FILTER. (B)(4).
A CUSTOMER REPORTED THAT THE ILLUMINATION SYSTEM TURNED OFF ON ITS OWN DURING A PROCEDURE. FOLLOWING A DELAY OF MORE THAN 15 MINUTES, THE PROCEDURE WAS COMPLETED WITH NO IMPACT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 139116 | ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM | ENDOILLUMINATOR | MPA | ALCON - IRVINE TECHNOLOGY CENTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |