FDA Adverse Event Malfunction Summary report: N

CD HORIZON® SPINAL SYSTEM

MDR report key: 3042812 · Received April 9, 2013

Report

Report Number
1030489-2013-00953
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 27, 2013
Report Date
May 15, 2013
Manufacturer
MEDTRONIC SOFAMOR DANEK PUERTO RICO OPERATIONS
Product Code
NKB
PMA / PMN Number
K113174
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION:ANALYSIS OF THE RETURNED DEVICE SHOWS THAT MACROSCOPIC AND OPTICAL EXAMINATION REVEALED THREAD CREST AND FLANK DAMAGE; THIS DAMAGE APPEARS TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE DAMAGED PORTION OF THE THREADS. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH MISALIGNMENT OF THE CONNECTOR AND SET SCREW THREADS DURING CONSTRUCT ASSEMBLY.

Additional Manufacturer Narrative · 1

(B)(4): NEITHER THE DEVICE NOR FILMS OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SPINAL PROCEDURE, DURING THE PROCEDURE IT WAS REPORTED THAT THE THREADS OF THE SET SCREW WERE SHEARING AND PREVENTED ADVANCEMENT OF THE SET SCREW INTO THE SCREW HEAD. NO FURTHER DETAILS ARE KNOWN AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
145290 CD HORIZON® SPINAL SYSTEM ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE NKB MEDTRONIC SOFAMOR DANEK PUERTO RICO OPERATIONS 5540030 H12K0610

Patients

Seq Age Sex Outcome Treatment
1 00073 YR