CD HORIZON® SPINAL SYSTEM
Report
- Report Number
- 1030489-2013-00952
- Event Type
- Malfunction
- Date Received
- April 9, 2013
- Date of Event
- March 27, 2013
- Report Date
- May 15, 2013
- Manufacturer
- MEDTRONIC SOFAMOR DANEK PUERTO RICO OPERATIONS
- Product Code
- NKB
- PMA / PMN Number
- K113174
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL INFORMATION: ANALYSIS OF THE RETURNED DEVICE SHOWS THAT MACROSCOPIC AND OPTICAL EXAMINATION REVEALED THREAD CREST AND FLANK DAMAGE; THIS DAMAGE APPEARS TO HAVE INITIATED AT THE START OF THE THREAD, AND IS CONSISTENT AROUND THE DAMAGED PORTION OF THE THREADS. THE ABOVE OBSERVATIONS ARE CONSISTENT WITH MISALIGNMENT OF THE CONNECTOR AND SET SCREW THREADS DURING CONSTRUCT ASSEMBLY.
(B)(4): NEITHER THE DEVICE NOR FILMS OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT.
IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN SPINAL PROCEDURE, DURING THE PROCEDURE IT WAS REPORTED THAT THE THREADS OF THE SET SCREW WERE SHEARING AND PREVENTED ADVANCEMENT OF THE SET SCREW INTO THE SCREW HEAD. NO FURTHER DETAILS ARE KNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 147138 | CD HORIZON® SPINAL SYSTEM | ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE | NKB | MEDTRONIC SOFAMOR DANEK PUERTO RICO OPERATIONS | 5540030 | H13A5583 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00073 YR |