FDA Adverse Event
Malfunction
Summary report: N
STENOSCOP
MDR report key: 3042636
·
Received April 2, 2013
Report
- Report Number
- 1720753-2013-04492
- Event Type
- Malfunction
- Date Received
- April 2, 2013
- Date of Event
- March 21, 2013
- Report Date
- April 2, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE VIDEO CABLE CONNECTORS WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT MAKE AN EXPOSURE, NO X-RAY. THIS RESULTED A TOTAL LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OR INJURY OR DEATH ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 134338 | STENOSCOP | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | STENOSCOP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |