FDA Adverse Event Malfunction Summary report: N

6600

MDR report key: 3042536 · Received April 2, 2013

Report

Report Number
1720753-2013-04552
Event Type
Malfunction
Date Received
April 2, 2013
Date of Event
March 21, 2013
Report Date
April 2, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYS AND REPLACED THE HARD DRIVE AND RELOADED SOFTWARE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYS WOULD NOT COMPLETE BOOT UP. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
134979 6600 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 6600

Patients

Seq Age Sex Outcome Treatment
1