FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3042528 · Received April 3, 2013

Report

Report Number
1720753-2013-04568
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 25, 2013
Report Date
April 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN INTERLOCK FAILURE ERROR MESSAGE AND FAILED TO BOOT UP. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136337 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1