FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3042498 · Received April 3, 2013

Report

Report Number
1720753-2013-04589
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 19, 2013
Report Date
April 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE HIGH VOLTAGE CABLE CONNECTORS WERE REGREASED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS DISPLAYED A COMMUNICATION ERROR MESSAGE AND LOCKED UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
137348 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1