FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3042487 · Received April 3, 2013

Report

Report Number
1720753-2013-04599
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 27, 2013
Report Date
April 3, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REP EVALUATED THE SYSTEM AND ADJUSTED THE BATTERY CHARGER. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED AN INTERMITTENT CHARGER FAILED ERROR AT BOOT UP. THIS ERROR WILL PREVENT THE SYSTEM FROM BOOTING UP, RESULTING IN A LOSS OF IMAGING FUNCTIONALITY. THIS COULD RESULT IN THE DELAY OF A PROCEDURE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136259 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment