FDA Adverse Event
Malfunction
Summary report: N
9600
MDR report key: 3042472
·
Received April 3, 2013
Report
- Report Number
- 1720753-2013-04603
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 20, 2013
- Report Date
- April 3, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYSTEM AND REPAIRED THE PUSH-BUTTON X-RAY SWITCH. THE SYSTEM OPERATES AS INTENDED. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE FSE NOTED THE FOOT SWITCH WAS STICKING AND PRODUCED UNCOMMANDED X-RAYS. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136298 | 9600 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9600 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |