FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3042257 · Received April 9, 2013

Report

Report Number
2531779-2013-03830
Event Type
Malfunction
Date Received
April 9, 2013
Report Date
March 14, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 DATE OF SUBMISSION 05/28/2013-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL KEY CONTACTS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE OK KEYPAD BUTTON WAS INTERMITTENTLY UNRESPONSIVE AND REQUIRED MULTIPLE HARD PRESSES TO ELICIT A RESPONSE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
145827 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 26 YR