FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3042218 · Received April 9, 2013

Report

Report Number
2531779-2013-03823
Event Type
Malfunction
Date Received
April 9, 2013
Report Date
March 12, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(4) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE UP ARROW AND DOWN ARROW KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
145226 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 75 YR