FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3041927 · Received April 8, 2013

Report

Report Number
3008382007-2013-07303
Event Type
Malfunction
Date Received
April 8, 2013
Report Date
March 13, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING OTHER MESSAGE-STRIP FILL ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143361 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3384718

Patients

Seq Age Sex Outcome Treatment
1